StatSpin Express 4 Horizontal Centrifuge Model # M510. Intended Use: Rapid Separation of plasma and serum from primary gel collection tubes.

This recall has been terminated (originally issued March 4, 2010).

Company
Statspin, Inc., Dba Iris Sample Processing
Recall Initiated
January 27, 2010
Posted
March 4, 2010
Terminated
November 13, 2019
Recall Number
Z-0807-2010
Quantity
1510 units
Firm Location
Westwood, MA
Official Source
View on FDA website ↗

Reason for Recall

Centrifuge shield micro-switch failed due to damage, and the unit opened, ejecting pieces.

Distribution

Worldwide Distribution -- USA, CANADA, AUSTRALIA, JORDAN, UK, BELGIUM, FRANCE, CZECH, RUSSIA, KUWAIT', NORWAY, GREECE, TURKEY, DENMARK, FINLAND, SEOUL, KOREA, and HONG KONG.

Lot / Code Info

Serial Numbers: 00100 through 001679.

Root Cause

Device Design

Action Taken

Iris Sample Processing notified distributors by an Urgent Device Recall letter advising of the recall and the defect. Users are to immediately quarantine the affected products and choose between the two options listed in regards to correcting the affected centrifuge. Customers may return the unit or an upgrade kit with instructions will be sent to upgrade the unit. Questions should be directed to Gail Magrath at 800-782-8774 ext. 6123.