StatSampler 200 uL, Li Heparin (micro blood collector) Ref: SS2H Packed in Catalog Numbers: SS2H (I vial), SS2H-06 (6 kits/case), CH2C-I0 (I0 vials/kit) Intended for collection and processing of capillary blood to prepare plasma/serum.

This recall has been terminated (originally issued August 23, 2008).

Company
Statspin, Inc., Dba Iris Sample Processing
Recall Initiated
June 20, 2008
Posted
August 23, 2008
Terminated
May 30, 2012
Recall Number
Z-2234-2008
Quantity
171 vials
Firm Location
Westwood, MA
Official Source
View on FDA website ↗

Reason for Recall

EDTA inadvertently mixed with lithium heparin solution contained in capillary tubes.

Distribution

Worldwide Distribution --- USA including states of CA, CO, FL, KY, KS, GA, NC, NJ, OR, PA, and countries of Canada, Hong Kong, UK.

Lot / Code Info

Lot number: 7296 Expiration Date: 12/09; Lot 8031 Expiration Date: 3/10

Root Cause

Mixed-up of materials/components

Action Taken

Iris Sample Processing notified customers by telephone followed by an Urgent: Product Recall letter dated 6/19/08. Users are requested to return inventory, to not use product from Lot Number 7296, and to repeat tests where the recalled tubes were used. The firm identified an additional lot : 8031 was effected and recalled the lot to two consignees.