Smiths Pneupac babyPAC portable ventilator, model 100. Designed for use by qualified medical caregivers, paramedics and other trained personnel, for hospital, emergency and transport ventilation of patients during respiratory distress or insufficiency.

Company
Smiths Medical Pm, Inc.
Recall Initiated
December 5, 2005
Posted
January 24, 2006
Terminated
December 26, 2006
Recall Number
Z-0436-06
Quantity
44
Firm Location
Waukesha, WI

Reason for Recall

A potential safety related problem associated with the Pnuepac babyPac Ventilator model 100 has been discovered in that the interface between the air input hose and the air input fitting may leak, and if tightened too much, the air input fitting may loosen when trying to remove the air hose.

Distribution

CA, CO, IN,MN, NY, PA, TN, TX

Lot / Code Info

0504244, 0505158, 0506211, 0506218, 0506242, 0506248, 0506254, 0506265, 0506283, 0506289, 0506321, 0506346, 0506358, 0507127, 0507139, 0507234, 0507235, 0507236, 0507248, 0507249, 0507265, 0508173, 0508174, 0508195, 0508196, 0508197, 504227, 504254, 505139, 505142, 505145, 505257, 505271, 505276, 505286, 506155, 506158, 506169, 506171, 506194, 506233, 506241, 506343, 507317

Root Cause

Other

Action Taken

Consignees were provided a Safety Action Bulletin, number 05-SAB05, dated 12/05/05. This requests that affected devices be quarantined, Smiths Medical be contacted and devices ultimately be returned to Smiths Medical Pm, Inc. Service Departmentfor repair.