Magnetic Resonance Imaging Systems under the following names: Visart Excelart Excelart P2 Excelart P3 Excelart SPIN Excelart VANTAGE

Company
Toshiba American Med Sys Inc
Recall Initiated
October 6, 2004
Posted
December 2, 2004
Terminated
September 2, 2005
Recall Number
Z-0270-05
Quantity
67
Firm Location
Tustin, CA

Reason for Recall

Software defect causes images to be rotated 180 degrees.

Distribution

Nationwide

Lot / Code Info

All serial numbers.

Root Cause

Other

Action Taken

Firm sent a letter to consignees October 6, 2004 describing the observed conditions causing the flaw to manifest itself. A workaround solution is also described in the letter. A software correction is under development and will be installed by a TAMS representative in the future.