MultiDiagnost 4 - x-ray system, multi functional/universal system for general use.

This recall has been terminated (originally issued February 27, 2003).

Company
Philips Medical Systems
Recall Initiated
January 10, 2003
Posted
February 27, 2003
Terminated
April 11, 2005
Recall Number
Z-0586-03
Quantity
22 units
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

INTERNAL SHORT CIRCUIT - MAY ALLOW FOR STAND AND TABLE TO MOVE SPONTANEOUSLY IN UPWARD OR DOWNWARD MOVEMENT

Distribution

The firm distributed to 22 hospitals/medical centers located in CA, FL, IA, IL, IN, KY, MN, MO, NC, ND, NY, OH, SC, TN, TX, VA, and WV.

Lot / Code Info

Serial numbers include (but are not limited to): 4584520996, 4594911196, 4595090397, 4595290697, 4594871196, 4594841196, 4595130197, 4615641197, 4595240797, 459500197, 4615511097, 4615581097, 4584570996, 4615320897, 459516497, 4595070197, 4584320996, 4584721096, 4615661197, 4594980197  Part numbers: 9896 000 58511, 9896 000 58512, 9896 000 58521, 9896 000 58522

Root Cause

Other

Action Taken

The firm issued a letter dated January 10, 2003 to their customers advising of the problem and that a respresentative would be contacting customers to arrange a time to visit to install new interface board. This is being conducted under FCO 70800007