MultiDiagnost 4 - x-ray system, multi functional/universal system for general use.
- Company
- Philips Medical Systems
- Recall Initiated
- January 10, 2003
- Posted
- February 27, 2003
- Terminated
- April 11, 2005
- Recall Number
- Z-0586-03
- Quantity
- 22 units
- Firm Location
- Bothell, WA
Reason for Recall
INTERNAL SHORT CIRCUIT - MAY ALLOW FOR STAND AND TABLE TO MOVE SPONTANEOUSLY IN UPWARD OR DOWNWARD MOVEMENT
Distribution
The firm distributed to 22 hospitals/medical centers located in CA, FL, IA, IL, IN, KY, MN, MO, NC, ND, NY, OH, SC, TN, TX, VA, and WV.
Lot / Code Info
Serial numbers include (but are not limited to): 4584520996, 4594911196, 4595090397, 4595290697, 4594871196, 4594841196, 4595130197, 4615641197, 4595240797, 459500197, 4615511097, 4615581097, 4584570996, 4615320897, 459516497, 4595070197, 4584320996, 4584721096, 4615661197, 4594980197 Part numbers: 9896 000 58511, 9896 000 58512, 9896 000 58521, 9896 000 58522
Root Cause
Other
Action Taken
The firm issued a letter dated January 10, 2003 to their customers advising of the problem and that a respresentative would be contacting customers to arrange a time to visit to install new interface board. This is being conducted under FCO 70800007