COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.

This recall has been terminated (originally issued August 15, 2012).

Company
Roche Molecular Systems, Inc.
Recall Initiated
February 1, 2012
Posted
August 15, 2012
Terminated
July 16, 2013
Recall Number
Z-2209-2012
Quantity
2,132 boxes
Firm Location
Branchburg, NJ
Official Source
View on FDA website ↗

Reason for Recall

Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce

Distribution

Nationwide Distribution (USA).

Lot / Code Info

Lot Numbers: 11049503, 11053503, 11054503, 11525503 and 11526503.

Root Cause

Nonconforming Material/Component

Action Taken

On 1/17/2012 an "Analyzer Bulletin" notification was sent to Roche affiliates and on 2/1/2012 an "Analyzer Bulletin" notification was sent to all affected US Consignees. The notification described the product issue and provided recommended actions. Customers were referred to Roche Technical Support at 800-526-1247 for questions.