Boston Scientific, Z Flex 270, Steerable Sheath, Sterile EO, Rx only, Distributed by Boston Scientific Corporation, One Boston Scientific Place, Natick, MA 01760-1537, Manufactured by Greathbatch Medical, 2300 Berkshire Lane North, Minneapolis, MN 55441. To facilitate transvenous introduction of diagnostic / therapeutic catheters into the vasculature and into the chambers of the heart.

This recall has been terminated (originally issued September 26, 2012).

Company
Greatbatch Medical
Recall Initiated
September 6, 2012
Posted
September 26, 2012
Terminated
January 22, 2015
Recall Number
Z-2459-2012
Quantity
787
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

During the investigation of four returned Z Flex 270 Steerable Sheaths that were used with cardiac cryoablation balloon catheters, the internal PTFE sheath liners were found to display varying degrees of damage that could pose an embolic risk to a patient.

Distribution

Distributed in MA.

Lot / Code Info

Q2112752, Q2110939, Q2130137, Q2131250, Q2130212, Q2150143, W2162515, W2163769, W2160626, W2175891, W2172722, W2184492, W2196366, W2184490, W2205843

Root Cause

Nonconforming Material/Component

Action Taken

Greatbatch Medical sent a "SAFETY ALERT-Z-FLEX-270 STEERABLE SHEATH" letter dated September 6, 2012 to the affected customer. The letter identified the product, problem, and actions to be taken by the customer. The letter advised the customer to communicate with their customers to assure the that the recalled sheath model is not used in conjunction with cryoablation catheters. Contact the firm at 763-951-8214 for questions regarding this issue.