Biosense Webster MobiCath Bi-Directional Guiding Sheath. Small Curve catalog no. D140010, Large Curve catalog no. D140011. The MobiCath Bi-Directional Guiding Sheath is a flexible tipped percutaneous catheter introducer (or guiding sheath) designed for gaining access to the coronary systems. The device features adjustable tip geometry through use of a rotating handle to deflect the catheter. This catheter is assembled with a hemostasis valve and a sideport infusion line. The MobiCath Bi-Directional Guiding Sheath is intended for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.

This recall has been terminated (originally issued December 22, 2012).

Company
Greatbatch Medical
Recall Initiated
November 12, 2012
Posted
December 22, 2012
Terminated
May 15, 2014
Recall Number
Z-0611-2013
Quantity
10,201 units
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Greatbatch Medical observed some anomalies of the inner lumen of the MobiCath Bi-Directional Guiding Sheaths upon implementation of a new visual inspection tool. The anomalies may include loose, string-like liner material, liner abrasion (fraying), and/or attached string-like liner material (from scratches or damage on the inner wall of sheath lumen).

Distribution

Worldwide Distribution-USA including the states of AL, AK, AZ, ALR, CA, CO, CT, DC, FL, GA, HI, IL IN, IA, KY, MD, MA, MI, MN, MO, NE, NV,NM, NJ, NH, NY, NC, OH, OK OR, PA, RI,SC,SD, TN, TX, UT,VT, VA, WA, and WI, and the countries of AUSTRIA, BELGIUM, SWITZERLAND, GERMANY, DENMARK, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, GREECE, HUNGARY, IRELAND, ISRAEL, ITALY, THE NETHERLAND, NORWAY, SWEDEN, AUSTRALIA, and CANADA

Lot / Code Info

LOT: R1701247 , R1713591 , W2046680 , W2046681 , W2072907 , W2078460 , W2084872 , W2098771 , W2102124 , W2102133 , W2102134 , W2108295 , W2108296 , W2108298 , W2108299 , W2125688 , W2125690 , W2132584 , W2132588 , W2132590 , W2138965 , W2143147 , W2143149 , W2143152 , W2145645 , W2145677 , W2145680 , W2145681 , W2151287 , W2151305 , W2151308 , W2156746 , W2156748 , W2156750 , W2156751 , W2162496 , W2162497 , W2162498 , W2162499 , W2167986 , W2167987 , W2167988 , W2167990 , W2172705 , W2172706 , W2172707 , W2172708 , W2172710 , W2172712 , W2177727 , W2177728 , W2177733 , W2177734 , W2184476 , W2184483 , W2184484 , W2184487 , W2190457 , W2190461 , W2190464 , W2190470 , W2196349 , W2196351 , W2196352 , W2196355 , W2196359 , W2196361 , W2196363 , W2211842 , W2211844 , W2211846 , W2211849 , W2211852 , W2211853 , W2213887 , W2229974 , W2236588.

Root Cause

Nonconforming Material/Component

Action Taken

Biosense Webster sent an "Urgent Field Safety Notice" dated November 12, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Voluntary Field Removal Certification Form was attached for customers to complete and return. Contact the firm at 800-729-9010 for questions related to this issue.