RecallDepth

Tecan clinical workstation; Tecan Genesis 150 front end, Roche catalog number 03585379001.

This recall has been terminated (originally issued July 20, 2004).

Company
Roche Diagnostics Corp.
Recall Initiated
March 5, 2004
Posted
July 20, 2004
Terminated
August 7, 2007
Recall Number
Z-0996-04
Quantity
14
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

If the files are not periodically purged from the TCW computer, it may result in patient identification/sample results mismatches.

Distribution

Nationwide. California, Connecticut, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, New Jersey, North Carolina, Ohio, Pennsylvania, Tennessee, Texas and Wisconsin.

Lot / Code Info

Serial numbers 5524, 6177, 6224, 6376, 6393, 6395, 6397, 6557, 7150, 7349, 7350, 7594, 7616 and 7724.

Root Cause

Other

Action Taken

Roche sent each of their consignees revised Tecan database maintenance instruction via bulletin 04-042 dated 3/5/04.

More recalls by Roche Diagnostics Corp.

View all recalls by this company →