Baxter Single Day Infusor 2 mL/h Portable Elastomeric Infusion System; an Rx sterile disposable drug delivery system; Baxter Healthcare Corporation, Deerfield, IL 60015; REF 2C1071KJP The Infusor device is indicated for patients requiring slow, continuous intravenous, intra-arterial, epidural, or subcutaneous administration of medications.

Company
Baxter Healthcare Corp. Rt
Recall Initiated
March 15, 2011
Posted
May 5, 2011
Terminated
March 12, 2012
Recall Number
Z-2146-2011
Quantity
3,696 units
Firm Location
Round Lake, IL

Reason for Recall

Some of the Single Day Infusor System 2 mL/h have a foil over-pouch that was incorrectly labeled as the Half Day Infusor System 5 mL/h, product code 2C1073KJP, lot 10H054.

Distribution

Worldwide Distribution - USA including AL and the countries of Mexico and Spain

Lot / Code Info

product code 2C1071KJP, lot 10H054 and product code 2C1073KJP, lot 10H054

Root Cause

Labeling mix-ups

Action Taken

Baxter Healthcare Corp. sent Urgent Product Recall letters dated March 15, 2011, via first class mail to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The accounts were requested to examine all of their inventory locations for both product code 2C1071KJP and 2C1073KJP, lot 10H054, and return the affected lot for replacement or credit. The accounts were also requested to complete the attached customer reply form listing the number of held for return, and fax it to Baxter at the number provided on the form. Any questions regarding the communication were directed to Baxter's Center for One Baxter at 1-800-422-9837.