Siemens brand 550 TxT Patient Table, Part No. 7346534, Distributed by and/or Manufactured by Siemens AG, Medical Solutions, Germany

This recall has been terminated (originally issued October 28, 2010).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
March 31, 2010
Posted
October 28, 2010
Terminated
March 1, 2012
Recall Number
Z-0166-2011
Quantity
326 units
Firm Location
Concord, CA
Official Source
View on FDA website ↗

Reason for Recall

An increase in voltage may cause a short circuit of a power transistor that may cause a smoldering and smoking effect of the electrical wiring and also an unexpected lowering of the Table. Both issues present the potential for injury to a patient or the operator.

Distribution

Worldwide Distribution including USA

Lot / Code Info

All codes.

Root Cause

Other

Action Taken

Siemens sent a Customer Information Letter on March 31, 2010, to all consignees. The letter identified the product, the problem, and the action that should be taken. Siemens implemented minor modifications to address the issues. No further action was required for the continual operation of the 550TxT Table. For questions regarding this recall call 925-602-8083.