RecallDepth

Boston Scientific ACCUSTICK II Introducer System with .038 inch Stainless Steel J Tip and .018 inch Nitinol Guidewire, sterile, REF/Catalog No. 20-710; UPN/Material Number: M001207100, Boston Scientific Corporation, Natick, MA 01760

This recall has been terminated (originally issued December 7, 2007).

Company
Boston Scientific Corp
Recall Initiated
July 19, 2007
Posted
December 7, 2007
Terminated
February 17, 2008
Recall Number
Z-0328-2008
Quantity
249 (199 U.S.A. plus 50 international)
Firm Location
Spencer, IN
Official Source
View on FDA website ↗

Reason for Recall

Marker Detachment: Out of specifications for pull strength, thus the marker may detach from the introducer during use.

Distribution

Worldwide: USA and The Netherlands.

Lot / Code Info

Lot /Batch Number: 9498023.

Root Cause

Process control

Action Taken

Consignees were notified via an Urgent Medical Device Recall Letter sent federal express on 7/19/07, which requested them to check their stocks, remove product from inventory, and return the product.