PRIMUS Plus,Medical charge-particle radiation therapy system (Medical Digital Linear Accelerator),MODEL NUMBER: 4504200,MANUFACTURED BY: Siemens Medical Solutions USA, Inc, 4040 Nelson Ave, Concord, CA 94520
This recall has been terminated (originally issued November 9, 2007).
- Recall Initiated
- March 22, 2007
- Posted
- November 9, 2007
- Terminated
- December 18, 2010
- Recall Number
- Z-0169-2008
- Quantity
- 687 units
- Firm Location
- Concord, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment plans. A second reason is an unplanned Dosimetry calibration data set can be erroneously copied during "Service Mode."
Distribution
Worldwide:USA including states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Delaware, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Vermont, Virginia, Washington, West Virgina, Wisconsin, and Wyoming and FOREIGN COUNTRIES: Canada, Mexico, Serbia Montenegro, Angola, Argentina, Australia, Austria, Belarus, Belgium, Bosnia, Herzeg-Bosnia, Brazil, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, DPR of Korea, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Malaysia, Morocco, Nepal, Netherlands, New Zealand, Norway, P.R. China, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uzbekistan, Venezuela, Vietnam, and Zambia
Lot / Code Info
MODEL NUMBER: 4504200; Serial Numbers: 2701, 2733, 2762, 2768, 2823, 2827, 2828, 2836, 2839, 2842, 2846, 2848, 2863, 2867, 2887, 2893, 2894, 2896, 2899, 2911, 2912, 2914, 2931, 2933, 2934, 2936, 2941, 2954, 2958, 2972, 2991, 2996, 2997, 3004, 3014, 3019, 3025, 3027, 3039, 3049, 3053, 3066, 3068, 3069, 3075, 3076, 3077, 3079, 3087, 3092, 3097, 3102, 3105, 3109, 3111, 3112, 3114, 3125, 3131, 3134, 3137, 3142, 3153, 3154, 3161, 3165, 3168, 3174, 3179, 3181, 3186, 3203, 3209, 3211, 3212, 3213, 3214, 3215, 3223, 3224, 3225, 3239, 3241, 3244, 3245, 3254, 3255, 3257, 3261, 3264, 3267, 3272, 3286, 3292, 3297, 3302, 3305, 3318, 3319, 3323, 3324, 3342, 3349, 3354, 3361, 3365, 3367, 3377, 3392, 3402, 3403, 3413, 3414, 3417, 3418, 3422, 3423, 3424, 3426, 3427, 3443, 3452, 3453, 3456, 3461, 3471, 3474, 3477, 3479, 3484, 3501, 3504, 3514, 3515, 3516, 3519, 3524, 3533, 3534, 3542, 3547, 3564, 3565, 3568, 3569, 3572, 3574, 3583, 3595, 3597, 3601, 3602, 3622, 3634, 3651, 3657, 3666, 3667, 3672, 3675, 3679, 3695, 3696, 3701, 3709, 3711, 3731, 3745, 3751, 3773, 3781, 3825, 3829, 3832, 3835, 3862, 3899, 3922, 3924, 3942, 3949, 3978, 3979, 4001, 4036, 4059, 5008, 50941, 2933A, 70-4117, 70-4185, 3744, 3749, 340258, 340262, 2956, 2963, 3579, 3582, 3587, 3592, 3663, 2773, 2815, 2964, 3046, 3217, 3308, 3334, 3724, 3917, 3598, 3752, 3373, 3561, 104744, 3042, 3158, 3191, 3284, 3291, 3295, 3356, 3378, 3382, 3406, 3407, 3469, 3619, 3638, 3641, 3645, 3742, 3919, 70-4173, 3612, 3824, 3995, 3732, 2869, 3219, 3343, 3502, 3861, 2871, 3268, 70-4126, 2753, 2808, 2928, 2987, 3026, 3032, 3037, 3047, 3117, 3201, 3234, 3346, 3398, 3431, 3522, 3545, 3599, 3649, 3703, 3704, 3767, 3814, 5063, 2782, 2783, 2796, 2829, 2832, 2833, 2851, 2854, 2866, 2891, 2938, 2942, 2944, 2945, 2948, 2967, 2973, 2989, 2994, 3031, 3033, 3034, 3044, 3072, 3081, 3091, 3093, 3094, 3098, 3106, 3127, 3128, 3129, 3144, 3145, 3155, 3182, 3199, 3231, 3259, 3332, 3333, 3348, 3351, 3362, 3387, 3399, 3438, 3467, 3539, 3549, 3552, 3578, 3614, 3686, 3693, 3763, 4013, 2957, 3307, 3395, 3444, 3936, 2862, 2976, 3642, 3833, 4099, 4131, 4232, 70-4131, 70-4308, 2793, 2805, 2918, 3202, 3499, 3571, 3577, 3684, 3714, 3718, 3725, 3768, 3786, 3794, 3958, 3972, 3977, 3999, 4042, 4075, 5073, 98943, 102842, 340274, 70-4264, 70-4194, 3311, 3369, 3401, 2785, 2992, 3107, 3135, 3183, 3326, 3372, 3396, 3408, 3416, 3421, 3475, 3689, 3961, 4039, 5036, 70-4133, 70-4271, ZTVK00010, 104, 2844, 2908, 3022, 3058, 3143, 3148, 3157, 3184, 3187, 3189, 3193, 3233, 3237, 3256, 3262, 3278, 3315, 3327, 3336, 3359, 3366, 3436, 3488, 3513, 3518, 3521, 3526, 3529, 3531, 3536, 3668, 3677, 3717, 3813, 3819, 3822, 3839, 3844, 3867, 3868, 3872, 3876, 3902, 3943, 3984, 3985, 3991, 4033, 4043, 4047, 4055, 4057, 4085, 4105, 5007, 5010, 5021, 5050, 5066, 5079, 5090, 70-4101, 70-4102, 70-4187, 70-4204, 70-4206, 70-4219, 70-4225, 70-4227, 70-4247, 70-4252, 70-4266, 70-4275, 70-4351, 70-4357, 70-4364, 3250, 4103, 70-4103, 3288, 2975, 3227, 3339, 3347, 3386, 3389, 3994, 4045, 70-4373, 3028, 3722, 4061, 70-4195, 344125, 2985, 3011, 3483, 3506, 3562, 3698, 3412, 3002, 3273, 3276, 3489, 3492, 2916, 2917, 2922, 2978, 2979, 3005, 3007, 3067, 3085, 3115, 3119, 3149, 3176, 3248, 3265, 3299, 3455, 3459, 3462, 3464, 3468, 3511, 3558, 3573, 3586, 3632, 3646, 3669, 3762, 3766, 3782, 3818, 3831, 3853, 3915, 3947, 3959, 3976, 3993, 4051, 4076, 4239, 5047, 340203, 70-4116, 70-4156, 70-4162, 70-4198, 70-4205, 70-4208, 70-4212, 70-4237, 70-4255, 70-4279, 70-4315, 70-4318, 70-4322, 70-4326, 70-4341, 70-4376, 3043, 3594, 340180, 3746, 340263, 3384, 3146, 3283, 70-4282, 2897, 2902, 2904, 2993, 3249, 3626, 3628, 3893, 4004, 3727, 3728, 3441, 70-4327, 2905, 3121, 3123, 3177, 3216, 3584, 3757, 3828, 3312, 3317, 2882, 3279, 3734, 4218, 70-4218, 2924, 3024, 3606, 3608, 3909, 70-4139, 70-4141, 70-4251, 2797, 2883, 2903, 2919, 2923, 2986, 3036, 3095, 3122, 3159, 3172, 3242, 3391, 3454, 3466, 3543, 3553, 3556, 3621, 3636, 3653, 3654, 3661, 3662, 3713, 3772, 3967, 3969, 4006, 4028, 4109, 5045, 5046, 70-4112, 70-4192, 70-4285, 2878, 3017, 3329, 2872, 2966, 3064, 3797, 3897, 5013, 2892, 3008, 3074, 3218, 3226, 3375, 3486, 3567, 3493, 3003, 3507, 3735, 3759, 4086, 5003, 340264, 70-4256, 3113, 3238, 3405, 3465, 3481, 3691, 3706, 3708, 3719, 3721, 340251, 298700097, 302600003, 303200005, 303700007, 304700009, 311700023, 334600031, 3411, 3887, 3525, 3544, 70-4302.
Root Cause
Software design
Action Taken
Distribution of Update Instructions TH007/07/S was sent to all affected sites on March 26, 2007. This was intended to inform users that Service Mode is only for use by authorized personnel, and that unauthorized and/or incorrect softpot parameter changes performed in Service Mode can result in the risk of patient mistreatment. Current control system software versions do not allow unplanned Dosimetry calibration data sets to be erroneously copied during "Service Mode."