M1020B PulseOximetry Module Philips FAST Sp02 and Pulse Oximetry Nellcor OxiMax Compatible used with the IntelliVue or M3/M4 patient monitor M1002

Company
Philips Medical Systems
Recall Initiated
September 21, 2006
Posted
November 16, 2006
Terminated
August 11, 2020
Recall Number
Z-0184-2007
Quantity
442 units
Firm Location
Andover, MA

Reason for Recall

Unexpected pulse oximetry (Sp02) readings (100%) over time when sensor is not attached to patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM

Distribution

Worldwide, including USA, Canada, Australia, China, Hong Kong, India, Indonesia, Japan, Phillipines, Singapore, Taiwan, Thailand, Vietnam, Argentina, Brazil, Chile, Colombia, El Salvador, Jamaica, Mexico, Nicaragua. Panama, Paraguay, Peru, Venezuela, Austria, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Germany, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Lithuania, Luxembourg, Macedonia, Mauritania, Mauritius, Netherlands, Norway, Oman, Slovenia, South Africa, Spain, Switzerland, Turkey, UAE, UK, Yemen, Zambia, Zimbabwe

Lot / Code Info

M1020B #AO01 AND #A02 with S/N DE524xxxxx M102B #A01 and #A02 with S/N DE612xxxxx With Sp02 Firmware (Fw) Rev: Sp02 DSP Fw-Rev. A.01.42

Root Cause

Other

Action Taken

Philips issued a Customer Letter and Urgent Device Correction letter to consignees on 09/21/06. Users were notified of the problem and provided with Instructions to Mitigate Risk. A software upgrade to: version D.00.61 and Sp02 FW A.01.46 is mandatory.