NanoSight LM10 fitted with LM14 module. A laboratory instrument for determination of particle characteristics (e.g. size).
- Recall Initiated
- February 28, 2022
- Posted
- September 26, 2022
- Recall Number
- Z-1762-2022
- Quantity
- 76 units
Reason for Recall
Product found to not comply to Class 1 Laser safety requirement.
Distribution
US Nationwide Distribution
Lot / Code Info
CDRH accession # 0510530-002
Root Cause
Radiation Control for Health and Safety Act
Action Taken
Customers are being made aware of this issue. Malvern Panalytical will retrofit modified microscope plates and cell tops with a new safety cover on the LM10 to maintain the original Class 1 laser product classification at no charge to the purchaser.