RecallDepth

NanoSight NS300

Company
Malvern Panalytical Ltd Grovewood Road Malvern United Kingdom
Recall Initiated
April 7, 2022
Recall Number
Z-1442-2022
Quantity
507 units

Reason for Recall

Product found to not comply to Class 1 Laser safety requirement.

Distribution

US Nationwide - Worldwide Distribution

Lot / Code Info

CDRH accession #0510530-003

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Customers are being made aware of this issue. All affected customers will be provided with a remediation kit with Class 3R labels and manual addenda with revised safety instructions.