RecallDepth

NanoSight NS300

This recall is currently active, issued April 7, 2022. It was issued by Malvern Panalytical Ltd Grovewood Road Malvern United Kingdom.

Company
Malvern Panalytical Ltd Grovewood Road Malvern United Kingdom
Recall Initiated
April 7, 2022
Recall Number
Z-1442-2022
Quantity
507 units
Official Source
View on FDA website ↗

Reason for Recall

Product found to not comply to Class 1 Laser safety requirement.

Distribution

US Nationwide - Worldwide Distribution

Lot / Code Info

CDRH accession #0510530-003

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Customers are being made aware of this issue. All affected customers will be provided with a remediation kit with Class 3R labels and manual addenda with revised safety instructions.