NanoSight NS300
- Recall Initiated
- April 7, 2022
- Recall Number
- Z-1442-2022
- Quantity
- 507 units
Reason for Recall
Product found to not comply to Class 1 Laser safety requirement.
Distribution
US Nationwide - Worldwide Distribution
Lot / Code Info
CDRH accession #0510530-003
Root Cause
Radiation Control for Health and Safety Act
Action Taken
Customers are being made aware of this issue. All affected customers will be provided with a remediation kit with Class 3R labels and manual addenda with revised safety instructions.