Olympus Resection Sheath, 28 Fr. with Deflecting Obturator. Model No. A22043T. Resection sheath for urologic applications.

Company
Olympus Corporation Of The Americas
Recall Initiated
November 11, 2025
Posted
February 20, 2026
Recall Number
Z-1446-2026
Quantity
207 units
Firm Location
Center Valley, PA

Reason for Recall

Complaints of the ceramic tip of the resection sheath breaking have been received.

Distribution

US Nationwide Distribution.

Lot / Code Info

Model No. A22043T; UDI: 04042761021012; All Lots.

Root Cause

Under Investigation by firm

Action Taken

An URGENT MEDICAL DEVICE CORRECTIVE ACTION notice was sent to consignees 1/6/2026. Consignees are instructed to ensure all personnel are familiar with the contents of the notification. Users are to adhere to the Inspection and Testing and Caution sections of the IFU for both the resection/inner sheath, as well as the Inspection Before Each Use section in the Resectoscope System Manual. Additional instructions include having an extra resection sheath and inspect during and after procedure for device fragments. Consignees are to retain a copy of the notification with the IFU. Consignees with any questions may contact Cynthia Ow at Cynthia.Ow@Olympus.com or by phone at 947-999-3203.