LithoCrushV Single Use Mechanical Lithotriptor V. Model Number: BML-V442QR-30. Single Use Mechanical Lithotriptor V used with Olympus endoscope for crushing calculi inside the bile duct.

This recall is currently active, issued January 16, 2025. It was issued by Olympus Corporation Of The Americas.

Company
Olympus Corporation Of The Americas
Recall Initiated
December 16, 2024
Posted
January 16, 2025
Recall Number
Z-0962-2025
Quantity
989 units
Firm Location
Center Valley, PA
Official Source
View on FDA website ↗

Reason for Recall

Potential for distal tip tearing of the Mechanical Lithotriptor V.

Distribution

US Nationwide distribution.

Lot / Code Info

UDI-DI: 04953170218422. Model Number: BML-V442QR-30. Lots: 33K-39K, 3XK, 3YK, 3ZK, 41K-44K

Root Cause

Under Investigation by firm

Action Taken

Olympus notified consignees on 12/16/2024, via URGENT: MEDICAL DEVICE REMOVAL letter. Consignees were instructed to examine inventory for affected devices, cease usage of affected devices, and arrange for return of any affected devices. Consignees were also instructed to acknowledge receipt of the letter through Olympus online portal as well as to notify customers if the affected devices were further distributed.