Medline Iris Scissors, Single-Use, Curved, Standard 4.5 , Model Nos. DYNJ04049 (cases), DYNJ04049H (eaches)
This recall is currently active, issued November 14, 2024. It was issued by Medline Industries, Lp - Northfield.
- Recall Initiated
- October 21, 2024
- Posted
- November 14, 2024
- Recall Number
- Z-0465-2025
- Quantity
- 69,611
- Firm Location
- Northfield, IL
- Official Source
- View on FDA website ↗
Reason for Recall
Tip protector may fall off within the packaging, resulting in compromised sterile barrier system and loss of sterility.
Distribution
US Nationwide distributions in the states of AL, AR, CO, CT, FL, IA, IN, KS, LA, MA, MO, NC, NY, OK, OR, RI, TN, TX, VA, WA, WY.
Lot / Code Info
UDI-DI: 10080196847273 Lots 23GBV681 23HBD138 23IBF254 23LBJ787
Root Cause
Under Investigation by firm
Action Taken
This is an expansion of recall RES 92881, which was originally initiated on April 30, 2024. Additional impacted customers were identified and notified on October 7, 2024. Customers were instructed to destroy any affected product on hand. The firm will issue credit for affected product.