RecallDepth

ENT, COEN27N; Medical convenience kit

This recall is currently active, issued September 24, 2024. It was issued by American Contract Systems, Inc..

Company
American Contract Systems, Inc.
Recall Initiated
August 7, 2024
Posted
September 24, 2024
Recall Number
Z-3241-2024
Quantity
66 units
Firm Location
Kansas City, MO
Official Source
View on FDA website ↗

Reason for Recall

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Distribution

US Nationwide distribution in the states of MO, MN, MA, OH, NE.

Lot / Code Info

UDI-DI: 00191072206244; Lot: 2405092; Exp: 05/09/2025

Root Cause

Under Investigation by firm

Action Taken

Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 08/07/2024 via email. The notice explained the issue, potential risk to health, and requested the following actions be taken: " -Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. - Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 08/07/2024 via email. The notice explained the issue, potential risk to health, and requested the following actions be taken: " Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. - Discard all affected kits and document on the enclosed Response Form."

More recalls by American Contract Systems, Inc.

View all recalls by this company →