Endovascular AAA Pack, ESED52B; Medical convenience kit

Company
American Contract Systems, Inc.
Recall Initiated
August 7, 2024
Posted
September 24, 2024
Recall Number
Z-3228-2024
Quantity
360 units
Firm Location
Kansas City, MO

Reason for Recall

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Distribution

US Nationwide distribution in the states of MO, MN, MA, OH, NE.

Lot / Code Info

UDI-DI: 00191072198402; Lot: 2405152; Exp: 05/15/2025

Root Cause

Under Investigation by firm

Action Taken

Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 08/07/2024 via email. The notice explained the issue, potential risk to health, and requested the following actions be taken: " -Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. - Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 08/07/2024 via email. The notice explained the issue, potential risk to health, and requested the following actions be taken: " Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. - Discard all affected kits and document on the enclosed Response Form."

More recalls by American Contract Systems, Inc.

View all recalls by this company →