TourniKwik Tourniquet Set (CFN 79012)
- Company
- Medtronic Perfusion Systems
- Recall Initiated
- May 16, 2024
- Posted
- July 23, 2024
- Recall Number
- Z-2423-2024
- Quantity
- 636 units
- Firm Location
- Brooklyn Park, MN
Reason for Recall
Incorrect component placed in four manufactured lots of the TourniKwik" Tourniquet Set.
Distribution
Worldwide - US Nationwide distribution in the states of MN, MA, UT, CA, and FL. The countries of Japan and Slovenia.
Lot / Code Info
Model Number: 79012; Serial Number: 2023020354/UDI: 20613994576358, Serial Number: 2023020355/UDI: 00673978186269, Serial Number: 202304C012/UDI: 20613994576358, and Serial Number: 202305C244/UDI: 20613994576358
Root Cause
Under Investigation by firm
Action Taken
Consignees were notified of the recall via mail beginning on 05/16/2024. Consignees were instructed to review inventory for listed product, immediately identify and quarantine all unused, listed product in inventory, return unused, listed product in the inventory to Medtronic, complete the customer confirmation form, and share the notification with orders in the organization.