URO ROBOTIC PACK , Model No LVUR44C LVUR44C-03

Company
American Contract Systems Inc
Recall Initiated
February 6, 2025
Posted
March 14, 2025
Recall Number
Z-1345-2025
Quantity
41 kits
Firm Location
Tiffin, OH

Reason for Recall

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Distribution

US Nationwide distribution in the state of OH.

Lot / Code Info

UDI-DI 00191072136688 Lots 905241 70-050739 70-051020 70-051480 70-051606 70-051888

Root Cause

Process control

Action Taken

The affected distributor was notified via email on 2/16/25, and an updated letter was provided on 2/18/25. The distributor was instructed to notify their end users. Customers were provided with a list of impacted components which are included in the affected trays. Customers were provided with a labeling template and instructed to add warning labels to affected procedure trays to ensure the affected components are discarded.