Medline Convenience kits, labeled as: 1) NERVE BLOCK TRAY-LF, Pack Number DYNJRA0377B; 2) PERIPHERAL NERVE BLOCK, Pack Number PAIN0137
This recall is currently active, issued September 6, 2024. It was issued by Medline Industries, Lp - Northfield.
- Recall Initiated
- April 8, 2024
- Posted
- September 6, 2024
- Recall Number
- Z-2988-2024
- Firm Location
- Northfield, IL
- Official Source
- View on FDA website ↗
Reason for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide distribution.
Lot / Code Info
DYNJRA0377B, Lot Number 19DDB791 ; DYNJRA0377B, Lot Number 21KDA359 ; PAIN0137, Lot Number 21CLA312; PAIN0137, Lot Number 21ALA608; PAIN0137, Lot Number 20LLA425; PAIN0137, Lot Number 20JLA143; PAIN0137, Lot Number 20HLA191; PAIN0137, Lot Number 20GLA104; PAIN0137, Lot Number 20CLA627; PAIN0137, Lot Number 20BLA634; PAIN0137, Lot Number 20ALA243; PAIN0137, Lot Number 21CLA312; PAIN0137, Lot Number 21ALA608; PAIN0137, Lot Number 20LLA425; PAIN0137, Lot Number 20JLA143; PAIN0137, Lot Number 20HLA191; PAIN0137, Lot Number 20GLA104; PAIN0137, Lot Number 20CLA627; PAIN0137, Lot Number 20BLA634; PAIN0137, Lot Number 20ALA243
Root Cause
Under Investigation by firm
Action Taken
Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.