Merge PACS

This recall has been terminated (originally issued August 6, 2020).

Company
Merge Healthcare, Inc.
Recall Initiated
August 6, 2020
Terminated
November 30, 2022
Recall Number
Z-2944-2020
Quantity
125 total units
Firm Location
Hartland, WI
Official Source
View on FDA website ↗

Reason for Recall

Measurements done on the Merge PACS generated MPR s may have incorrect measurements.

Distribution

Domestic Distribution: IL, MO, NY, ND, MI, TX, FL, OH, UT, NE, MS, MN, DE, AR, CA, WI, MD, NC, NH, GA, VA, SC, PE, CT, NM, IN. International Distribution: CA, BE, UK.

Lot / Code Info

Unique Device Identifier (UDI): v8.1:  PACS:(01)00842000100805(10)8.1(11)191022 OrthoPACS: (01)00842000100805(10)8.1(11)191022 v81.1:  PACS: (01)00842000100805(10)8.1.1(11)200528 OrthoPACS: (01)00842000100805(10)8.1.1(11)200528  V8.1.2  PACS: (01)00842000100805(10)8.1.2(11)200709

Root Cause

Software design

Action Taken

The firm, IBM Watson Health, sent an "Urgent Field Safety Notice" dated 05 August 2020 to its customers. The letter described the product, problem, and actions to be taken. The customers were instructed to do the following: Avoid incorrect pixel measurements by using the following workflow: Step 1: Modality acquires a series and generates reformat and sends to Merge PACS Step 2: Radiologist performs measurement on these series from the modality. Avoid black dots that may lead to additional imaging by taking any of the following actions: - Verifying study images either at modality or in another viewer. - Increasing the window level will window out the lowest pixel value (the black dots). - Viewing additional images/series within the study. Viewing of prior series/studies. - Verifying patients history, diagnosis or other clinical data. IBM Watson Health/Merge Healthcare will provide a software correction to address these issues. A representative will be in contact with you to schedule the software update. Please ensure all potential users of Merge PACS v8.1, 8.1.1 and 8.1.2 have received this notice. Please provide this notice to other organizations impacted by this action. If you have any questions regarding this notice, please contact Customer Support at WHISupport@us.ibm.com or 877-741-5369.