LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).

Company
Smith & Nephew Inc
Recall Initiated
December 16, 2019
Recall Number
Z-0878-2020
Quantity
59 units
Firm Location
Memphis, TN

Reason for Recall

Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.

Distribution

USA: IA, NC, PA, TN. Int'l: CH, BE, GB, AE, CN, DE, NL, BR, FR.

Lot / Code Info

LEGION AP Cutting Block Size 4, Product Number 71434409, Lot Numbers: 17LM05399, 17LM07852, 17LM07852A; LEGION AP Cutting Block Size 6, Product Number 71434411, Lot Numbers: 17LM05397, 17LM05398, 17LM07849; LEGION AP Cutting Block Size 7, Product Number 71434412, Lot Numbers: 17LM01746, 17LM05390, 17LM05394, 17LM07851; LEGION AP Cutting Block Size 8, Product Number 71434413, Lot Number: 17LM08442.

Root Cause

Process change control

Action Taken

The firm initiated the recall by letter on December 16, 2019. The letter identified the problem and directed the consignee to remove the affected lots noted in the Urgent Medical Device Recall Notice.