Hillrom Welch Allyn Spot Vision Screener VS100, marketed under the following models: a) SPOT VISION SCREENER,W/O CASE,PLUG2/EUR, VS100-2; b) SPOT VISION SCREENER,W/O CASE,PLUG4/UK, VS100-4; c) SPOT VISION SCREENER,W/O CASE,PLUG7/SA, VS100-7; d) SPOT VISION SCREENER,W/CASE,PLUG2/EUR, VS100S-2; e) SPOT VISION SCREENER,W/CASE,PLUG4/UK, VS100S-4; f) SPOT VISION SCREENER,W/CASE,PLUG7/SA, VS100S-7; g) SPOT VISION SCREENER W/CASE, BRAZIL, VS100S-Z; h) SPOT VS POWER SUPPLY SET/PLUG2/EUR, 106364; i) SPOT VS POWER SUPPLY SET/PLUG4/UK, 106366; j) SPOT VS POWER SUPPLY SET/PLUG7/SA, 106368; for use during vision screening

This recall is currently active, issued April 10, 2024. It was issued by Baxter Healthcare Corporation.

Company
Baxter Healthcare Corporation
Recall Initiated
March 26, 2024
Posted
April 10, 2024
Recall Number
Z-1510-2024
Quantity
3547 units
Firm Location
Deerfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.

Distribution

Global, US: NY, IL

Lot / Code Info

a) VS100-2, UDI/DI 732094224214; b) VS100-4, UDI/DI 732094224191; c) VS100-7, UDI/DI 732094224092; d) VS100S-2, UDI/DI 732094224177; e) VS100S-4, UDI/DI 732094224153; f) VS100S-7, UDI/DI 732094224085; g) VS100S-Z, UDI/DI 732094294217; h) 106364, UDI/DI 732094246964; i) 106366, UDI/DI 732094319927; j) 106368, UDI/DI 732094248692

Root Cause

Nonconforming Material/Component

Action Taken

Baxter issued an Urgent Medical Device Correction notice to its consignees on 03/26/2024 via USPS, first class mail. The notice explained the issue, hazard, and provided the following: Actions to be taken by customers: 1. Healthcare providers may continue to use the non-compliant power cords, but if any fraying or damage is observed, users should NOT make physical contact with the cord and discard the power cord immediately. 2. Healthcare providers should regularly inspect the power cords for fraying or other damage. 3. Once Baxter has replacement power cords, a follow-up notification will be sent with additional instructions on how to request replacement power cords. 4. Customers will be asked to acknowledge receipt of the Urgent Medical Device Correction notification by responding on our customer portal. 5. If the product has been distributed to other facilities or departments within the institution, customers will be asked to forward a copy of the Urgent Medical Device Correction notification to them. 6. If the customer is a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, they will be asked to notify their customers of this Urgent Medical Device Correction according to their customary procedures and check the associated box on the customer portal.