RecallDepth

Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

This recall is currently active, issued January 29, 2024. It was issued by Xtant Medical Holdings, Inc.

Company
Xtant Medical Holdings, Inc
Recall Initiated
November 28, 2023
Posted
January 29, 2024
Recall Number
Z-0846-2024
Quantity
6
Firm Location
Belgrade, MT
Official Source
View on FDA website ↗

Reason for Recall

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

Distribution

US distribution in MI.

Lot / Code Info

UDI-DI: 00810076564005, Lot: AP23000143

Root Cause

Process control

Action Taken

On 11/28/23 the recall notice was emailed. Customers with questions can call the firm at 1-406-924-5878 from 8 a.m. to 5 p.m. (Mountain Time).

More recalls by Xtant Medical Holdings, Inc