Cortera Spinal Fixation System Screw, Poly, Solid, 6.5mm x 45mm, REF: 150100-65045

This recall is currently active, issued October 17, 2024. It was issued by Xtant Medical Holdings, Inc.

Company
Xtant Medical Holdings, Inc
Recall Initiated
August 21, 2024
Posted
October 17, 2024
Recall Number
Z-0125-2025
Quantity
330
Firm Location
Belgrade, MT
Official Source
View on FDA website ↗

Reason for Recall

Spinal fixation system implant screws marked "DNI" are not intended to be implanted and may not have been manufactured in accordance with released manufacturing specifications, and if implanted the screws may not have adequate strength or performance.

Distribution

US Nationwide distribution in the states of IL, WI, FL, PA, CA, TX, MI, MO, VA, OH, MA.

Lot / Code Info

UDI-DI: 00810076563190, Lot: DNI

Root Cause

Process change control

Action Taken

On 8/21/24, recall notices were mailed and emailed to customers who were asked to do the following: 1) Examine affected devices and quarantine any screws with DNI markings. 2) Notify your staff about the recall. If affected devices were further distributed, provide the recall notice to these customers. 3) Complete and return the response forms via email to mdolan@xtantmedical.com