QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-10X3-E, Order Number G22667; used to dilate strictures of the biliary tree

This recall is currently active, issued December 1, 2023. It was issued by Wilson-Cook Medical Inc.

Company
Wilson-Cook Medical Inc
Recall Initiated
November 3, 2023
Posted
December 1, 2023
Recall Number
Z-0458-2024
Quantity
111 devices
Firm Location
Winston Salem, NC
Official Source
View on FDA website ↗

Reason for Recall

Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.

Distribution

US

Lot / Code Info

UDI/DI 00827002227675 (each), 10827002227672 (case), Lot Numbers: W4739833, W4749234, W4750262, W4756463, W4756465, W4720843, W4722160, W4722641, W4723307, W4724321, W4739835, W4740109, W4740421, W4746519, W4747434, W4748249, W4749510, W4749744, W4750867, W4751104, W4751904, W4752276, W4752565, W4752902, W4752908, W4755364, W4756899, W4757275, W4758645, W4759509, W4760713, W4761742, W4761743

Root Cause

Under Investigation by firm

Action Taken

Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice on 11/03/2023 by mail. The notice explained the problem with the device, risk to health, and requested the following actions be taken: Customer: Cease use, quarantine, and return all affected devices. If devices were further distributed, those to whom the devices were sent must be notified. Distributor: Quarantine all affected devices, notify your customers to whom you distributed devices. Return all devices to Cook Medical.