Centricity PACS-IW with Universal Viewer, versions 5.0 SPx with PACS-IW foundation; Radiological Image Processing System
- Company
- Ge Healthcare
- Recall Initiated
- September 14, 2023
- Posted
- October 20, 2023
- Recall Number
- Z-0139-2024
- Quantity
- 17 units
- Firm Location
- Chicago, IL
Reason for Recall
GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and Centricity Universal Viewer V6.0 with PACS-IW foundation.
Distribution
US and Austria, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, Chile, Colombia, France, Germany, India, Ireland, Italy, Kuwait, Malaysia, Netherlands, Netherlands Antilles, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Kingdom, Uruguay, Venezuela, Vietnam
Lot / Code Info
UDI/DI none, Installed Product ID #: 100707-1-Centricity Universal Viewer-00558649, 1045-1-Centricity Universal Viewer-00558651, 1045-1-Centricity Universal Viewer-19607785, 100759-1-Centricity Universal Viewer-02360986, 9842-4-Centricity Universal Viewer-19386293, 4723-1-Centricity Universal Viewer-02303576, 104648-2-Centricity Universal Viewer-02354806, 101272-1-Centricity Universal Viewer-00558671, 100713-1-Centricity Universal Viewer-00558259, 100713-1-Centricity Universal Viewer-01982629, 4705-1-Centricity Universal Viewer-02398923, 100512-1-Centricity Universal Viewer-00558264, 3727-1-Centricity Universal Viewer-00558421, HC3062-Centricity Universal Viewer-01462397, RU9761-UW PACS-IW UP TO 10K-02336112, ZA2533-Centricity Universal Viewer-00116577, YP3150-Centricity Universal Viewer-01639018
Root Cause
Software design
Action Taken
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/14/2023 by letter. The notice explained the problem, risk to the patient, and provided two workaround options for the continued use of the instruments until the software update is ready for installation. Questions: please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative