RecallDepth

Centricity PACS-IW with Universal Viewer, versions 5.0 SPx with PACS-IW foundation; Radiological Image Processing System

Company
Ge Healthcare
Recall Initiated
September 14, 2023
Posted
October 20, 2023
Recall Number
Z-0139-2024
Quantity
17 units
Firm Location
Chicago, IL

Reason for Recall

GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and Centricity Universal Viewer V6.0 with PACS-IW foundation.

Distribution

US and Austria, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, Chile, Colombia, France, Germany, India, Ireland, Italy, Kuwait, Malaysia, Netherlands, Netherlands Antilles, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Kingdom, Uruguay, Venezuela, Vietnam

Lot / Code Info

UDI/DI none, Installed Product ID #: 100707-1-Centricity Universal Viewer-00558649, 1045-1-Centricity Universal Viewer-00558651, 1045-1-Centricity Universal Viewer-19607785, 100759-1-Centricity Universal Viewer-02360986, 9842-4-Centricity Universal Viewer-19386293, 4723-1-Centricity Universal Viewer-02303576, 104648-2-Centricity Universal Viewer-02354806, 101272-1-Centricity Universal Viewer-00558671, 100713-1-Centricity Universal Viewer-00558259, 100713-1-Centricity Universal Viewer-01982629, 4705-1-Centricity Universal Viewer-02398923, 100512-1-Centricity Universal Viewer-00558264, 3727-1-Centricity Universal Viewer-00558421, HC3062-Centricity Universal Viewer-01462397, RU9761-UW PACS-IW UP TO 10K-02336112, ZA2533-Centricity Universal Viewer-00116577, YP3150-Centricity Universal Viewer-01639018

Root Cause

Software design

Action Taken

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/14/2023 by letter. The notice explained the problem, risk to the patient, and provided two workaround options for the continued use of the instruments until the software update is ready for installation. Questions: please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative