GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.
- Company
- Ge Healthcare
- Recall Initiated
- February 19, 2025
- Posted
- March 28, 2025
- Recall Number
- Z-1487-2025
- Quantity
- 4 units
- Firm Location
- Chicago, IL
Reason for Recall
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
Distribution
Worldwide distribution.
Lot / Code Info
Model/catalog/code # 2049587-015, Installed Product ID 100707-1-Centricity Universal Viewer-00558649, UDI UDI Not Required, System ID PACS-IW-126; Model/catalog/code # 2066908-086, Installed Product ID 9842-4-Centricity Universal Viewer-19386293, UDI UDI Not Required, System ID GA001PACTRL01; Model/catalog/code # 2066908-136, Installed Product ID ZA2533-Centricity Universal Viewer-00116577, UDI UDI Not Required, System ID ZA2533PAC1; Model/catalog/code # 2104867-045, Installed Product ID X9003062 - HC3062-Centricity Universal Viewer-01462397, UDI (01)00840682103800(10)6.0SP50412094097001D, System ID HC3062UV01.
Root Cause
Software design
Action Taken
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 by letter delivered using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your device. Please ensure all potential users in your facility are made aware of this safety notification. Please ensure you have implemented the actions below: 1: Ensure your hospital network is secure and 2: Outside of the hospital network implement additional safeguards such as a VPN. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.85480@gehealthcare.com." "If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative." Per status report 05/23/2025, the firm send a follow-up notice dated 05/13/2025 providing an installation manual addendum that provides instructions for changing passwords.