CELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood Part Number: 7900001

Company
Menarini Silicon Biosystems
Recall Initiated
May 20, 2022
Recall Number
Z-1345-2022
Quantity
484 units
Firm Location
Huntingdon Valley, PA

Reason for Recall

High number of total images/unassigned events including (dual positives) and the potential for false positive results being placed into the image gallery in some patient samples

Distribution

CA FL IL IN MA MI NC NM NY OH ON PA TX Foreign: Italy, Shanghai, Singapore,

Lot / Code Info

UDI-DI: (01)10888385000016(17)221004(10)S108 Lot Number: S108 Exp. Date: 04Oct2022

Root Cause

Process control

Action Taken

Menarini Silicon Biosystems issued notification telephone Urgent Medical Device Correction Notification CTC Kit Lot S108 to US on 5/20/22. Followed by e-mail direct to consignees for US sites - May 23, 2022 and Coordination with Asia Pacific region and the EU with Local QA and RA from the regions. Letter states reason for recall, health risk and action to take: Discontinue usage of Lot S108 and quarantine the lot in your inventory for disposal. Review patient samples run with Kit Lot S108 which had results that were above the cut-off. If you observe any images with the staining pattern of the false positive images shown below or there is any uncertainty regarding these images impacting the patient's outcome from below the cutoff to above the cut-off, the patient sample should be repeated. Report any previous occurrence of unusually high number of total images/ unassigned events or potential false positive events that have not already been reported to the Customer Technical Support. " Complete and return the enclosed Confirmation of Receipt form no later than May 31, 2022. If you have additional questions about this issue, contact Customer Technical Support at +1-877-837-4339 option 2

More recalls by Menarini Silicon Biosystems

View all recalls by this company →