RecallDepth

CELLSEARCH CIRCULATING Tumor Cell Kits (IVD) (Product Code 7900001) CELLSEARCH CIRCULATING Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of the epithelial origin (CD45-, EpCAM+, cytokeratins 8, 18+, and/or 19+) in whole blood.

Company
Menarini Silicon Biosystems
Recall Initiated
April 4, 2018
Terminated
April 9, 2019
Recall Number
Z-1530-2018
Firm Location
Huntingdon Valley, PA

Reason for Recall

Unusually high number of total images/unassigned events may require additional time and effort for the operator to review all the images and may cause a delay in reporting the results.

Distribution

Worldwide Distribution - USA (nationwide) and to the countries of : Canada and Italy

Lot / Code Info

UDI (01)10888385000012(17)180810(10)S038 (01)10888385000012(17)180810(10) (01)10888385000012(17)180810(10)S038S (01)10888385000012(17)180810(10)S038T (01)10888385000012(17)180810(10)S038U (01)10888385000012(17)180810(10)S038 V  Serial, Batch, Lot # 2038 S038R S038S S038T S038U S038V S038W (0 distributed to the field)

Root Cause

Process design

Action Taken

Menarini silicon biosystems sent an Urgent Medical Device Correction Notification dated April 3, 2018. Required Actions. " Please dispose of any affected product in your current inventory . " Report any previous occurrence of unusually high number of total images I unassigned events in patient samples to Customer Technical Support. " Complete and return the enclosed Confirmation of Receipt form no later than April 20,2018. For further question, please call (215) 830-0777.