Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO; and ECMO Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO

Company
Origen Biomedical, Inc.
Recall Initiated
August 21, 2018
Terminated
October 5, 2021
Recall Number
Z-0102-2019
Quantity
106
Firm Location
Austin, TX

Reason for Recall

The catheter failed the endotoxin testing.

Distribution

US Nationwide in the states of NY, AR, MO, NC, GA, LA, CO, AL, MA, CA, MI, TX, OH, MD, FL, TN, SC, OR, AZ, UT, RI, SD, PA, NM, CT. Foreign (OUS): United Arab Emirates, Kingdom of Saudi Arabia, New Zealand, Thailand

Lot / Code Info

UDI: 00816203022059, lot N18687-1

Root Cause

Release of Material/Component prior to receiving test results

Action Taken

On 08/21/18, the firm emailed a recall notice to customers informing them of a nationwide recall of RDLC ECMO catheters. Customers were advised to return affected product to the firm. Customers with additional questions were advised to call 1-800-233-9014 or 1-512-474-7278. The firm also sent customers Technical Bulletin (TB) 1411 - Revision Feb 2018, Subject: Radiopacity of the Reinforced Dual Lumen Catheter (RDLC), Date: July 2015, and an Acknowledgment Receipt form to complete and return. On 09/12/18, and 09/20/18, the firm sent two additional recall notices, which included additional affected lot numbers that were not the subject of the current removal.