cobas e 801 immunoassay analyzer

Company
Roche Diagnostics Corporation
Recall Initiated
March 15, 2019
Terminated
April 23, 2020
Recall Number
Z-1249-2019
Quantity
149
Firm Location
Indianapolis, IN

Reason for Recall

The firm has confirmed complaints regarding signal drops on the cobas e 801 module after a ProCell II M bottle changeover, which may lead to incorrect medical decisions with respect to diagnostics and patient treatment.

Distribution

Distributed to accounts in AL AZ CA CT FL HI IA IL IN KY LA MA MD ME MI MN MO NC NH NJ NY OH OR PA PR SC TN TX VA WA WI

Lot / Code Info

All units may experience this issue.

Root Cause

Device Design

Action Taken

On March 15, 2019, the firm notified affected customers of the recall via an Urgent Medical Device Correction letter. Customers were informed of the potential problem of signal drops when the standby bottle flowpath is covered with a biofilm. Customers were asked to do the following: " Follow the Workaround section to determine if your cobas e 801 module is potentially affected and perform the steps if applicable. " If your cobas e 801 module is potentially affected based on the information in the Workaround section of this UMDC, contact the Roche Support Network Customer Support Center at 1-800-428-2336 to schedule the ProCell II M flowpath decontamination procedure. The ProCell II M flowpath decontamination procedure should be performed every 4 weeks until your cobas e 801 module is switched to the improved ProCell II M formulation. " Identify and retest potentially affected patient samples using the Retrospectively Identify and Correct Potentially Affected Patient Sample Results section of this UMDC. " Complete the attached fax form and fax it to 1-833-254-2597 or email it to roche10068@stericycle.com.

More recalls by Roche Diagnostics Corporation

View all recalls by this company →