RecallDepth

TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail

This recall is currently active, issued August 22, 2023. It was issued by Smith & Nephew Inc.

Company
Smith & Nephew Inc
Recall Initiated
July 24, 2023
Posted
August 22, 2023
Recall Number
Z-2444-2023
Quantity
32 units
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

A complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT as described on the product label.

Distribution

US, AE, CA, DE, ES, GB, MX, PA, PL

Lot / Code Info

UDI/DI 03596010500922, Batch Number 21KSM0606

Root Cause

Process control

Action Taken

Smith & Nephew issued an Urgent Medical Device Recall Notice on 07/24/2023. The notice explained the problem with the product, the potential risk, and requested the consignees locate and quarantine the affected product for return. Sales reps, district offices, or distributors were directed to notify their customers and ensure that the required actions were complete. The firm is seeking return of the affected product.