cobas b 221<6>Roche OMNI S6 system; Catalog numbers: 3337154001 and 3337154692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
This recall has been terminated (originally issued May 3, 2018).
- Company
- Roche Diagnostics Corporation
- Recall Initiated
- February 22, 2018
- Posted
- May 3, 2018
- Terminated
- November 15, 2019
- Recall Number
- Z-1612-2018
- Quantity
- 1026 total products
- Firm Location
- Indianapolis, IN
- Official Source
- View on FDA website ↗
Reason for Recall
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Distribution
USA ( nationwide)
Lot / Code Info
Catalog numbers: 3337154001 and 3337154692 UDI: 04015630018345
Root Cause
Software design
Action Taken
ROCHE sent an Urgent Medical Device Correction Letter dated February 22, 2018, via UPS to affected customers. Customers were instructed to complete the following: " 1) Use the Return to Setup and Exit buttons to ensure execution of scheduled AutoQC measurements as described in this Urgent Medical Device Correction (UMDC); 2) Check to see if the next scheduled AutoQC measurement is performed to ensure the AutoQC scheduler is working as expected; 3) If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site; 4) Complete the attached fax form (TP-00146) and fax it to 1-844-491-7776 or scan and email to roche7570@stericycle.com; 5) File this UMDC for future reference. Customers with questions should contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272.