IMPAX, PACS, Picture Archive and Communications System, IMPAX CV 7.8.x and IMPAX CV 12.x. Corrects certain demographic information (Patient Name, Patient Medical Record Number and/or Accession Number in previously signed reports that have since changed.
This recall has been terminated (originally issued September 1, 2015).
- Company
- Agfa Healthcare Corp.
- Recall Initiated
- September 1, 2015
- Terminated
- December 15, 2020
- Recall Number
- Z-0283-2016
- Quantity
- 19
- Firm Location
- Greenville, SC
- Official Source
- View on FDA website ↗
Reason for Recall
Customers have experienced IMPAX CV software, specifically, Demographics Manager (DM), not producing correct pdf reports after the reports, at the time of physician signature, were correct.
Distribution
Distributed in the states of CA, KY, NH, NY, NC, OH, OR, PA, SC, SD, TN, TX, WA and the country of Canada.
Lot / Code Info
Model # L9M2100
Root Cause
Software design
Action Taken
An Urgent Field Safety Notice was sent to accounts via email on 9/1/2015. The letter described the safety alert. Customers were informed of the problem and that Service would be contacting them to arrange a date to deploy the corrective action and install the Service Update at their facility. An acknowledgement was included which was to be faxed back or emailed back to Agfa.