Agfa Healthcare NX 3.0.8950 Imaging Processing Software

This recall has been terminated (originally issued December 21, 2016).

Company
Agfa Healthcare Corp.
Recall Initiated
December 21, 2016
Terminated
July 10, 2019
Recall Number
Z-2735-2017
Quantity
221 units
Firm Location
Greenville, SC
Official Source
View on FDA website ↗

Reason for Recall

A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong image was linked to the exam and appeared. After a short time the wrong image was replaced by the correct image, however the wrong image was used for transmitting to PACS.

Distribution

AL, AZ, AR, CA, DC, FL, GA, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WV and Canada

Lot / Code Info

2UA15205XC, 2UA161JNH, A5411002007, A5411001003, A5411001002, 5CG5164040, 2UA5481BVY, 5CG5171X81, A5411002026, A5411002025, A5410000039, 2UA62524PS, 2UA62524Q2, 2UA62524RK, A5411002031, 2UA61218G4, 5CG61266X6, 2UA61218GG, 2UA63825HQ, 2UA642306H, 2UA4161JMY, 2UA52524Y5, 2UA62524RJ, 2UA62524QV, 2UA62524QB, 2UA62524RJ, 2UA62524NS, 2UA62524PJ, 2UA62524Q9, A5411000155, A5411000151, 2UA62524PH, 2UA62524QL, 2UA62524QT, A5411000153, 2UA62524Q0, 5CG6126B9J, 5CG61266YW, 5CG61266VP, 5CG61268BG, 5CG612673N, 2UA6243GNY, 2UA62424R8, 2UA62524P4, 2UA62524RY, 2UA6243GPT, 2UA62524Q6, 2UA62524PR, A5411000152, 2UA63825HR, 2UA31000Q6, 2UA31000Q8, 2UA31000R5, 2UA31000QR, 2UA319120W, 2UA3191211, 2UA31000PS, 2UA31000QX, 2UA31000Q0, 2UA31000PX, A5411000346, A5411000232, A541000199, A5411000210, A5411000209, 2UA62524P8, 5CG61266R5, 5CG5295760, 5CG5285W67, 5C5285W7P, 5CG5285W9Y, 5CG5285W5R, 5CG528W7B, A5411001163, A5411000468, 2UA52524XM, 2UA5481BTP, A5411000145, 2UA25009HM, A5411001064, A5411000159, 2UA3380VMW, 2UA3380VMG, 2UA3461QSD, A5411001039, A5411001066, A5411001065, A5411000230, A5411000177, A5411000174, CZC6349KSZ, 2UA62524RH, 2UA62524R4, 2UA6254Q7, A5411000116, A5411002095, A5411002093, A5411002092, A5411001173, A5411001172, 2UA6243GPK, 2UA6243GPL, 2UA62524QH, 2UA4150QFZ, 2UA4150QFQ, 2UA1110ZYG, 2UA3021KYP, 2UA52524XC, 2UA52524YK, 2UA52524YQ, A5411001105, MXL24220J8, A5411002030, A5411001089, A5411002058, 2UA2181HSV, A5411000238, 2UA6021JXG, 2UA511BZ2, A5411000341, A5411000340, A5411000337, 2UA5111C0V, CZ6357GKR, CZC6357GFM, A5411002051, A5411002048, A5411002047, A5411002052, A5411002049, A5411000549, 2UA62524RW, 2UA62524QS, 2UA62524P7, 2UA62524R7, 2UA62524QM, A5410001000, A5410000132, 2UA5170ZKM, 2UA5422QN4, 2UA62524PP, 2UA62524R6, 2UA1221B32, 2UA62524QD, A5411002033, 2UA62524RX, 5CG6126743, 2UA62524PB, 2UA44616DJ, 2UA63825J3, 2UA62524P6, 2UA62524QF, 2UA62524PL, 5CG61266QW, 2UA62524PN, 2UA62524RS, 2UA62524RQ, 2UA62524QJ, 5C61266Y4, A5411002064, A54110002059, CZC6349KXK, 2UA5111BZC, 2UA642306K, 2UA63825J4, 2UA64029N9, 2UA62524PD, 2UA62524P1, 2UA62524Q8, 2UA63825J6, 2UA63825JD, 2UA62524P0, 2UA63825HS, 2UA63825HN, 2UA63825J5, 2UA63825J8, 2UA61225X9, 2UA62524PT, 2UA63825J0, 2UA62524PV, 2UA5111BYZ, 5CG61266YZ, 2UA62524NY, 2UA62524RV, 5CG61266MG, 2UA63825JC, 2UA63825HX, 2UA62524QZ, 2UA62524QR, 2UA62524PF, 5CG53032V5, 2UA52223Q2, 2UA52223Q4, 2UA51519JR, 2UA62524RP, 2UA32878M, 2UA5451JKP, 2UA62524RG, 2UA62524R5, 2UA62524RN, 2UA63825HM, A5411002022, A45411000018, A5411002057, 2UA63825HZ, A5411000180, 2UA15205X9, 2UA62524NX, 2UA31000Q9, A5411000179, 2UA5451JL2, 2UA1510H6B, 2UA5311CS3, 2UA61218GT, CZC6057ZSL, A5411002024, 2UA62524RM, 2UA62524PY, 2UA1510H60, 2UA1510H5C

Root Cause

Device Design

Action Taken

On December 22, 2016, an URGENT FIELD SAFETY NOTICE letter was sent either by mail or email. The letter included an acknowledgement form to be returned. The letter informed the customer of the problem and actions to be taken.