BVI Custom Eyes Ophthalmic Procedure Kits with the component -Devon Light Glove Part Number: 584649 Ophthalmic surgical kits
This recall has been terminated (originally issued June 15, 2015).
- Recall Initiated
- May 1, 2015
- Posted
- June 15, 2015
- Terminated
- January 28, 2021
- Recall Number
- Z-1804-2015
- Quantity
- 585 kits
- Firm Location
- Waltham, MA
- Official Source
- View on FDA website ↗
Reason for Recall
Devon Light Glove in procedure kits may have holes/slits compromising sterility of the product
Distribution
Nationwide Distribution including KY, GA, MI, and NJ.
Lot / Code Info
Lot Numbers: 3039720 3046211 3047770 3049473 3070763 3082330 3089827 3093682 3095941 3106407 3117874
Root Cause
Nonconforming Material/Component
Action Taken
Beaver-Visitec International Sales Ltd. (BVI) sent an Urgent Field Safety Notice dated May 2, 2015, to all affected customers. Customers were instructed to quarantine the affected kits and return to BVI for subsequent replacement of product. A Verification Form was provided to confirm receipt of the recall notice. For questions regarding this recall call 791-906-7950.