bvi Beaver Xstar Safety Crescent, 2.5mm, 55o Bevel Up, REF 378234 Product Usage: Specialty blades and knives are used for stab and cut incisions in ophthalmic, ENT and other surgical applications.

Company
Beaver-Visitec International Inc.
Recall Initiated
November 3, 2017
Posted
February 1, 2018
Terminated
May 31, 2023
Recall Number
Z-0503-2018
Quantity
123,240 units
Firm Location
Waltham, MA

Reason for Recall

The protective shield was retracted, and the blade was exposed.

Distribution

Worldwide Distribution - US Nationwide and the countries of Italy, UK, Finland, Germany, Ireland, Turkey

Lot / Code Info

Batch Numbers: 3180336 3181395 3183925 3184461 3183518 3187183 3187940 3189616 3190463 3192521 3193360 3194101 3199801 3200589 3201368 3201959 3202091 3202764 3203228 3204381 3205075 3205511 3205828 3207499 3208102 3211276 3215327 3218819 3218144 3220074 3221091 3221092 3223818 3223055 3223115 3224020 3225234

Root Cause

Process control

Action Taken

The firm disseminated their safety advisory notices on 11/03/2017 by email and visit. The letter identified the affected product, problem and actions to be taken. The firm requested the following action:"Prior to opening the Tyvek tray, please visually confirm that the plastic shield is not retracted and the activation slider is fully engaged."