cobas 4800 KRAS AMP/DET 24T CE-IVD Mutation Test; CE-IVD 5852170190. Intended for the identification of mutations in codons 12,13 and 61 of the KRAS Gene in DNA derived from formalin-fixed paraffin-embedded human colorectal and non-small cell lung cancer tissues.

Company
Roche Molecular Systems, Inc.
Recall Initiated
December 18, 2014
Posted
June 26, 2015
Terminated
February 8, 2017
Recall Number
Z-0999-2015
Quantity
660 pieces
Firm Location
Branchburg, NJ

Reason for Recall

cobas KRAS Mutation Kit T10786 is generating invalid results for the Mutant Control and/or KRAS Calibrator.

Distribution

Foreign Distributed.

Lot / Code Info

Lots T10786 T13091, T15435, T15475, T14777, and W00389

Root Cause

Device Design

Action Taken

RMS issued an Urgent Field Safety Notice on 12/18/2014. A Quality Notification was provided to all users of the of cobas KRAS Mutation Kit CE-IVD, lots T13091, T15435, T15475, T14777, and W00389, informing them of the issue and instructing them to discontinue use and discard any remaining inventory immediately.