SHAPE ARM, SINGLE, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682400253 and the Arm part number in the kit is P17085. . The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS.

This recall has been terminated (originally issued July 15, 2014).

Company
Stryker Communications
Recall Initiated
March 17, 2014
Posted
July 15, 2014
Terminated
May 15, 2015
Recall Number
Z-2052-2014
Quantity
953 STRYKER SHAPE ARMS of all types
Firm Location
Flower Mound, TX
Official Source
View on FDA website ↗

Reason for Recall

During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.

Distribution

Worldwide Distribution. US nationwide including Puerto Rico, China, Philippines, Mexico, Korea, Japan, Colombia, Chile, Canada, Brazil, and Australia

Lot / Code Info

Serial numbers 111018DE001 through 140210DE020  Serial and part numbers are located on underside of the cover of the portion of the Arm that is attached to the boom's service head. Screws must be removed and the cover detached to view these numbers.

Root Cause

Process control

Action Taken

Direct accounts in the US were notified by letter dated March 21, 2014, and asked to report any malfunctions to Stryker Communications immediately. In addition, Stryker Representatives will visit all US accounts to inspect all the product units on hand. If needed, Representatives will check that black, screw/lock washers supplied by Stryker are secured in place and will replace any that are not the typed furnished by Stryker. All units will be updated with a yellow, warning sticker on the monitor bracket which reads: "Failure to properly install the screw and lock washer below can result in the attached component/equipment falling." Stryker Communications' foreign distributors will oversee notification, inspection, and corrections of their accounts and report corrections and other findings to Stryker Communications.