S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good Model Number o P38198 - ASM, MFR CONTROLLER, FIXED STANDARD o P38199 - ASM, MFR CONTROLLER, LIFT STANDARD o P40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD o P40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD o P46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD
- Company
- Stryker Communications
- Recall Initiated
- January 29, 2021
- Recall Number
- Z-1297-2021
- Quantity
- 1,393 units
- Firm Location
- Flower Mound, TX
Reason for Recall
The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY. The country of Canada.
Lot / Code Info
Serial Numbers: 2258046 2863490
Root Cause
Under Investigation by firm
Action Taken
An Urgent Customer Notification Letter was mailed (certified) to the affected consignees on 01/29/2021. Customers will be notified by letter and a Stryker Representative will visit each account for product inspection and replacement. If customers notice any issues prior to inspection, customers should discontinue use and call Technical Support immediately at (866) 841-5663.