Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination.

This recall has been terminated (originally issued June 7, 2014).

Company
Leica Microsystems, Inc.
Recall Initiated
September 11, 2013
Posted
June 7, 2014
Terminated
April 24, 2017
Recall Number
Z-1726-2014
Quantity
103 total
Firm Location
Buffalo Grove, IL
Official Source
View on FDA website ↗

Reason for Recall

The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).

Distribution

Worldwide Distribution -- USA, including the states of AZ, IL, MA, MO, NC, OR, TX, VA, and WI; and, the countries of Australia, Austria, Chile, China, Colombia, Czech Republic, Greece, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Oman, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, and United Kingdom.

Lot / Code Info

Leica Ceiling Mounts MS-1F, MC-1F; Article Numbers: 10446657, 10446658

Root Cause

Nonconforming Material/Component

Action Taken

A Medical Device Correction letter, dated September 11, 2013, was sent to all direct accounts. The letters advised customers that a Leica Microsystems representative will contact them and make arrangements to conduct an inspection of the ceiling mount(s) and to provide additional preventative maintenance and inspection protocols for the devices. Customers are asked to complete and return the enclosed Acknowledgement Form and send it back to Leica Microsystems within 10 days of receiving the Medical Device Correction letter. Customers with questions should contact their Leica Microsystems representative.