Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W. Sterile using ethylene oxide. Product Usage: The Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewires are intended to facilitate catheter placement and exchange during diagnostic or interventional procedures in the aorta, where increased support, distal flexibility, and low surface friction of the guidewire is needed.

This recall has been terminated (originally issued March 28, 2013).

Company
Medtronic Inc. Cardiac Rhythm Disease Management
Recall Initiated
March 13, 2013
Posted
March 28, 2013
Terminated
February 6, 2014
Recall Number
Z-1019-2013
Quantity
27,682 devices
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Medtronic has identified an issue involving specific lot numbers of the Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W, where a change in the manufacturing process has been identified as the cause for potential kinking and breaking at the proximal weld end of the outer spring coil during in-vivo use.

Distribution

Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of: Argentina, Armenia, Australia, Austria, Belgium, Bolivia, Canada, Chile, China, Colombia, Croatia, Czech Republic, Ecuador, El Salvador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Israel, Italy, Jordan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Venezuela, Viet Nam.

Lot / Code Info

Model ARCSJ200W Lot Numbers: GFWC1124, GFWC1506, GFWC2138, GFWC2652, GFWC2653, GFWD0353, GFWD0881, GFWD0882, GFWD2048, GFWD2049, GFWD2050, GFWD2051, GFWD2052, GFWE0743, GFWE0744, GFWE0745, GFWE0746, GFWE0747, GFWE3710, GFWE4272, GFWE4273, GFWE4274, GFWE4275, GFWF0143, GFWF0144, GFWF0147, GFWF0148, GFWF0149, GFWG0428, GFWG0429, GFWG0430, GFWG0431, GFWG0432, GFWG0433, GFWG0434, GFWH0333, GFWH0334, GFWH0335, GFWH0336, GFWH0337, GFWH0338, GFWI0190, GFWI0191, GFWI0192, GFWI0193, GFWI0194, GFWI0195, GFWJ0027, GFWJ0028, GFWJ0029, GFWJ2454, GFWJ2455, GFWJ2456, GFWJ2457, GFWJ2458, GFWK1910, GFWK1911, GFWL0370, GFWL2571, GFWL2572, GFXA0224.   Model ARCSJ260W Lot Numbers: GFWC1501, GFWC1502, GFWC1503, GFWD0883, GFWD0884, GFWD3256, GFWE0752, GFWE0753, GFWE3711, GFWF0145, GFWF0150, GFWG0425, GFWG0426, GFWG0427, GFWH0344, GFWH0345, GFWI0201, GFWI0202, GFWI0203, GFWJ0418, GFWJ0419, GFWJ0420, GFWJ2461, GFWJ2462, GFWJ2464, GFWJ2465, GFWL0368, GFWL0369, GFWL0415, GFWL0416, GFXA0219, GFXA0220, GFXA0221, GFXA0222.   Model ARCDJ260W Lot Numbers: GFWC1507, GFWC1508, GFWD0885, GFWE0750, GFWE0751, GFWF0142, GFWF0146, GFWG0423, GFWG0424, GFWH0331, GFWH0332, GFWI0186, GFWI0187, GFWI0188, GFWJ2449, GFWJ2450, GFWJ2451, GFWJ2452, GFWK1844.

Root Cause

Component change control

Action Taken

Medtronic sent an Urgent Medical Device Recall letter dated March 2013 to all affected customers. The letter identifies the affected product, problem and actions to be taken. Customers were instructed to immediately quarantine all affected products and returned to Medtronic. Customers were requested to complete the attached Customer Confirmation Certificate and fax it to Medtronic to the attention of Customer Focused Quality. For question contact your local Medtronic representative or Medtronic Customer Service at 1-888-283-7868.