RecallDepth

Kimberly-Clark MIC PEG Kit - Pull Type The Kimberly-Clark MIC PEG Kits - Pull type are Percutaneous Endoscopic gastrostomy (PEG) tubes intended for a non-surgical "pull" endoscopic placement technique. The PEG tube is held in place during use with an external retention disk.

This recall has been terminated (originally issued September 20, 2012).

Company
Kimberly-Clark Corporation
Recall Initiated
June 22, 2010
Posted
September 20, 2012
Terminated
February 7, 2013
Recall Number
Z-2416-2012
Quantity
766 units
Firm Location
Roswell, GA
Official Source
View on FDA website ↗

Reason for Recall

The placement guidewire diameter is too large and does not pass-through the introducer needle during the placement procedure of the PEG Feeding Tube.

Distribution

Nationwide Distribution including AL, AK, AZ, CA, CO, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, ME, MA, MI, MN, MS, MO, NE, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VA, WA, and WI.

Lot / Code Info

Lots AW916002A, AW916102D

Root Cause

Nonconforming Material/Component

Action Taken

Kimberly-Clark Health Care sent an Urgent Voluntary Medical Device Withdrawal letter dated June 22, 2010, to all affected customers. The letter included a Product Withdrawal Response Sheet which was to be returned to Kimberly-Clark. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and destroy all affected products and complete the attached Product Withdrawal Response Sheet and Fax the form to 920-225-4182. For questions customers should call 770-587-7362. For questions regarding this recall call 770-587-8393.