OPMI Pentero, OPMI Pentero C, OPMI Pentero 900, OPMI Vario/S8, OPMI Vario/S88, OPMI Neuro Multivision/NC4, OPMI Vario/NC33, OPMI Neuro/NC4, OPMI Neuro/NC4 Ceiling Mount. A surgical microscope and accessories is an AC powered device intended for use during surgery to provide a magnified view of the surgical area.

Company
Carl Zeiss Meditec Inc
Recall Initiated
September 12, 2011
Posted
September 23, 2011
Terminated
February 24, 2012
Recall Number
Z-3275-2011
Quantity
1677 consignees
Firm Location
Dublin, CA

Reason for Recall

Some third party surgical drapes have optically active protective lenses that could affect the ZEISS surgical microscope position information that is passed on to third party navigation systems.

Distribution

Nationwide distribution: USA only.

Lot / Code Info

All-OPMI¿ Pentero¿, OPMI¿ Pentero¿ C, OPMI¿ Pentero¿ 900, OPMI¿ Vario/S8, OPMI¿ Vario/S88, OPMI¿ Neuro Multivision/NC4,  OPMI¿ Vario/NC33, OPMI¿ Neuro/NC4, OPMI¿ Neuro/NC4 Ceiling Mount.

Root Cause

Pending

Action Taken

The firm, Zeiss, sent an "IMPORTANT SAFETY ALERT FOR ZEISS SURGICAL MICROSCOPES" letter dated September 7, 2011, to all affected customers. The letter described the product, problem and action to be taken. The customers were instructed to use surgical drapes compatible with ZEISS surgical microscopes and always follow the instructions for use of the third party navigation system manufacturer for calibration, calibration verification and operation. In addition, the firm supplied customers with a supplement to their Zeiss surgical microscope user documentation and requested that they read the information and keep a copy with their original instructions also the customers were instructed to complete and return the Acknowledgement of Receipt form within 5 business days via fax to 1-866-809-6044 or USPS to Stericycle, Inc., Attn: Event 2480, 2670 Executive Dr. Suite A, Indianapolis, IN 46241. Should you have any questions about this Safety Alert, please contact your Carl Zeiss Customer Care representative at 1-888-871-7120.