i-STAT PT/INR Cartridge Abbott Point of Care, Princeton, NJ

This recall has been terminated (originally issued July 13, 2010).

Company
Abbott Point Of Care Inc.
Recall Initiated
May 5, 2010
Posted
July 13, 2010
Terminated
September 22, 2011
Recall Number
Z-1996-2010
Quantity
44976 in US, 3648 in rest of world.
Firm Location
Princeton, NJ
Official Source
View on FDA website ↗

Reason for Recall

A high mean bias (12-16%) was documented in 8 lots of i-STAT PT/INR cartridges. This may lead a clinician to believe that a patient is adequately anticoagulated when they are not. Inadequate anticoagluation therapy may be given and issues with clotting and/or thrombosis could occur. No reports of injury have been received.

Distribution

Worldwide distribution: Argentina, Hong Kong, China, Finland, Greece, India, Portugal, Russian Federation, Saudi Arabia, United Arab Emirates, Italy, Spain, Australia, Germany, UK, Sweden, The Netherlands, France, Canada.

Lot / Code Info

Catalog numbers: 04J50-01, 04J50-02. Lot numbers: N09315, exp 4/28/2010; N09323, exp 5/14/2010; N09323A, exp 5/14/2010; N09346A, exp 5/28/2010; S09347, exp 5/28/2010; S09354, exp 6/14/2010; T10009, exp 6/28/2010; T10011, exp 6/28/2010.

Root Cause

Other

Action Taken

Urgent recall notifications were sent on May 13 2010. Customers were instructed to discontinue use and to return the identified lots. Additional questions are directed to Abbott Point of Care Technical Support at 800-3668020, Option 1 or your Abbott Point of Care representative.